Core Laboratory
2. C-Peptide Elecsys Reagent Kit, Cat. No. 03184897 - 100 tests
The electrochemiluminescence
immunoassay “ECLIA” is intended for use on the Roche Elecsys
1010/2010 and MODULAR ANALYTICS E170 (Elecsys module) immunoassay analyzers.
Principle:
Sample, a biotinylated monoclonal C-peptide-specific antibody, and a monoclonal C-peptide-specific antibody labeled with a ruthenium complex (Tris(2,2’-bipyridyl)ruthenium(II)-complex (Ru(bpy)2+3 ) react to form a sandwich complex. After addition of streptavidin-coated microparticles, the complex becomes bound to the solid phase via interaction of biotin and streptavidin. The reaction mixture is aspirated into the measuring cell where the microparticles are magnetically captured onto the surface of the electrode. Unbound substances are then removed with ProCell. Application of a voltage to the electrode then induces chemiluminescent emission which is measured by a photomultiplier. Results are determined via a calibration curve which is instrument-specifically generated by 2-point calibration and a master curve provided via the reagent barcode.
Specimen Collection and
Preparation:
Only the specimens listed below were tested and found acceptable.
1.
Serum collected using standard sampling tubes or tubes containing
separating gel.
2.
Li-heparin and K3-EDTA plasma.
Criterion: Recovery within 90-110% of serum value or slope 0.9-1.1+
coefficient of correlation > 0.95.
Assay within 2-3 hours, or store frozen at –20oC
for 1-2 weeks.
Measuring Range:
3 - 13300 pmol/L or 0.01-40 ng/mL (defined by the lower
detection limit and the maximum of the master curve).
Values below the detection limit are reported as < 3 pmol/L (< 0.01 ng/mL).
Values above the measuring range are reported as > 13300 pmol/L (> 40 ng/mL)
Reference Ranges:
370 – 1470 pmol/L
Precision:
Serum and plasma:
Reproducibility was determined using Elecsys reagents, pooled human
sera, and controls in a modified protocol (EP5-A) of the NCCLS (National
Committee for Clinical Laboratory Standards): 6 times daily for 10 days
(n = 60); human serum 3 at one day 5 times (n = 59); within-run
precision on MODULAR ANALYTICS E170 analyzer (n = 21). The following
results were obtained:
MODULAR ANALYTICS E170
E170
Within-run precision
Total precision
|
Sample |
Mean pmol/L |
SD |
CV |
Mean |
SD |
CV |
|
MC 3 |
330 |
2 |
0.6 |
310 |
6 |
1.9 |
|
Human serum 2 |
640 |
3 |
0.5 |
610 |
10 |
1.6 |
|
Human serum 3 |
2000 |
19 |
0.9 |
1920 |
44 |
2.3 |
|
Human serum 4 |
5990 |
54 |
0.9 |
|
|
|
|
Human serum 5 |
8590 |
126 |
1.5 |
|
|
|
|
PreciControl MA1 |
700 |
6 |
0.8 |
680 |
11 |
1.6 |
|
PreciControl MA2 |
3600 |
31 |
0.9 |
3390 |
55 |
1.6 |
Analytical Sensitivity (lower detection limit):
3 pmol/L (0.01 ng/mL)
The detection limit represents the lowest measurable analyte level that
can be distinguished from zero. It is calculated as the value lying two
standard deviations above that of the lowest standard (master
calibrator, standard 1 + 2 SD, within-run precision, n = 21).
Method Comparison:
Serum
A comparison of the Elecsys C-Peptide assay (y) with a commercially
available C-peptide assay (x) using clinical serum samples gave the
following correlations:
Number of samples measured: 266
Passing/Bablok
Linear regression
y = 1.07x + 0.03
y = 1.11x - 0.15
τ = 0.962
r = 0.996
SD (md68) = 0.117
Sy.x = 0.242
The sample concentrations were between approx. 157 and 7260 pmol/L or
approx. 0.470 and 21.8 ng/mL.
Analytical Specificity:
For the monoclonal antibodies used, the following cross-reactivities were found:
|
Substance |
Concentration tested |
Cross-reactivity |
|
Proinsulin, humane |
0.10 |
32.5 |
|
Insulin, human |
8.66 |
0.005 |
|
Insulin, porcine |
7.50 |
not detected |
|
Insulin, bovine |
7.69 |
not detected |
|
Somatomedin(IGF-I) |
1.0 |
not detected |
|
Human Growth Hormone |
10.0 |
not detected |
|
Glucagon |
10.0 |
not detected |
The Elecsys C-Peptide assay uses two monoclonal antibodies specifically directed against human C-peptide. The antibodies show cross-reactivity with the C-chain of human proinsulin and presumably with partially processed proinsulins (split products). The concentrations of proinsulin and split products of fasting healthy subjects are 100 times lower than the C-peptide concentrations and therefore the cross-reactivity is of no clinical significance. In patients with insulinoma, the proinsulin concentrations are reported as up to 60-fold higher than those from fasting healthy subjects.
For more information, please contact the Core Laboratory manager or director.